For the first time, a team of researchers at the Technical University of Munich (TUM) in Germany has integrated the dark-field x-ray method into a CT scanner suitable for clinical use. Dark-field imaging provides additional information to conventional x-ray imaging. With the new prototype, it is now possible to produce 3D dark-field x-ray images.

“For the first time, we showed that dark-field x-ray technology can also be integrated into a clinical CT scanner,” says Franz Pfeiffer, PhD, who headed the study. Although this technology is in its early stages, pre-clinical studies with mice have demonstrated clear benefits from dark-field CT scans, especially for capturing images of lung tissue.”

The new CT prototype has already been used successfully with a thorax phantom, a model of a human upper body, and is large enough for the intended applications with real patients.

Conventional X-ray Imaging

With conventional x-ray equipment, the x-rays are attenuated by the intervening tissue as they travel from the source to the detector. This effect is used to produce images based on the varying degrees of attenuation associated with different tissue types and structures. That is why bones and similar structures, which have a stronger attenuating effect, appear white in X-rays, while more transparent tissue types such as the lung produce darker images.

Dark-field X-ray Imaging

Dark-field imaging, by contrast, makes use of the small-angle scattering of the x-rays. When the x-rays interact with materials of different densities such as the interface between lung tissue and air, they are scattered. The analysis of this scattering effect yields additional information on very fine tissue structures, which is otherwise not accessible with conventional x-ray images.

Grating technology for Dark-Field Imaging

To detect the scattering of the x-ray radiation, a set of three optical gratings is required. They are placed between the x-ray source and detector. When x-rays pass through these gratings, a characteristic pattern is produced at the detector. When a sample or person is then positioned in the beam path, this characteristic pattern is changed. These deviations are then used to analyze the structure of the sample or the person’s tissue.

New Hardware and Software for Dark-Field CT 

The implementation of the dark-field method in a human-size CT scanner poses various technical challenges. Until now, this has limited dark-field CT devices to a scale much smaller than would be needed for human patients. Apart from the size, the fast rotation of the scan unit also creates special difficulties for the technical design.

The scanning unit of CT scanners, known as the gantry, rotates at very high speeds. This causes vibrations that affect the finely tuned components in the interior of the device. Based on a detailed analysis of these vibrations, the team was able to use them to implement the required shift between the gratings needed for dark-field imaging. To analyze the scans, they developed new algorithms to filter out the vibration effects based on reference scans.

Additional Information for Clinical Diagnostics

“With the dark-field CT prototype, we can capture conventional and dark-field x-ray images in a single scan. This yields additional information that could be used in the future not only to diagnose lung diseases, but also to differentiate between various types of kidney stones and tissue deposits,” says Manuel Viermetz, MSc, one of the two first authors of the study. As the next step, the researchers plan to further optimize the dark-field CT prototype and prepare for the first scans of human patients.